The frequency of diagnostic ultrasound is inversely proportional to depth of penetration.
The Minidop has 5 interchangeable probes with different frequencies.
Use those probes depending on your applications.
BT2M20S8C (2 MHz): Deep peripheral blood flow
BT4M05S8C (4 MHz):Deep peripheral blood flow
BT5M05S8C (5 MHz):Deep peripheral blood flow
BT8M05S8C (8 MHz):Superficial blood flow
BT10M5S8C (10MHz):Superficial blood flow


Clinical applications
Detection of Arterial and Venous Blood Flow Velocity for vascular disease.


Probes to be used:
BT2M20S8C (2 MHz)
BT4M05S8C (4 MHz)
BT5M05S8C (5 MHz)
BT8M05S8C (8 MHz)
BT10M5S8C (10MHz)


Probe selection
Standard Doppler probe:
Standard:
2MHz:BT2M20S8C
4MHz:BT4M05S8C
5MHz:BT5M05S8C
8MHz:BT8M05S8C
10MHz:BT10M5S8C
Pencil:
8MHz:BP8M05S8A
10MHz:BP10M5S8A
Flat:
2MHz:BF2M20S8A
8MHz:BF8M15S8A
Sterilizable probes: (Amplifier required)
Small pencil:
8MHz:VRP-08
10MHz:VRP-10
20MHz:VRP-20


Sterilization
Only sterilizable probes can be sterilized. Do not sterilize other type of probes
including amplifiers as well as main unit.
Sterilization limits
All sterilizable probes except ACP and FDP probe: Up to 50 times
FDP probe: Up to 5 times
ACP probe: Up to 5 times (steam autoclave)
Note: Do notexceed sterilization limits or it may cause damage to probes.
Caution
Except ACP probe, do not sterilize probes by steam autoclave nor put them in washer disinfector or it will damage probes.
ACP probe should be sterilized by steam autoclave as described in section "Sterilization" below.
Cleaning:
Automated
Do not do automated cleaning of probes other than ACP probe.
Manual
Do not soak it into medicinal solution. Wipe any contamination from probes with damp cloth.